DocumentCode
3687010
Title
A novel calibration procedure of laser induced plasma spectroscopy for quantitative bulk analyses of archaeological bronzes
Author
J. Agresti;S. Siano
Author_Institution
Ist. di Fis. Appl. “
fYear
2015
fDate
5/1/2015 12:00:00 AM
Firstpage
1
Lastpage
4
Abstract
Here, the effectiveness of a new calibration model for the characterization of archaeological bronze artefacts using laser induced plasma spectroscopy is demonstrated. Deep depth profiles are mandatory in the analysis of archaeological bronzes where intentional patinations and corrosion phenomena may significantly have altered the composition of their outer layers up to several hundreds of micrometers. The work shows that the standard calibration curves approach may produce errors in the quantitative analysis, which are corrected using 2D calibration surfaces.
Publisher
iet
Conference_Titel
Fotonica AEIT Italian Conference on Photonics Technologies, 2015
Print_ISBN
978-1-78561-068-4
Type
conf
DOI
10.1049/cp.2015.0173
Filename
7322083