• DocumentCode
    3689188
  • Title

    Surface head coil for 10.5 tesla magnetic resonance imaging (MRI)

  • Author

    Elizaveta Motovilova;Shao Ying Huang;Jiasheng Su

  • Author_Institution
    Singapore University of Technology and Design 8 Somapah Road, Singapore 487372
  • fYear
    2015
  • fDate
    7/1/2015 12:00:00 AM
  • Firstpage
    1
  • Lastpage
    3
  • Abstract
    In this paper, we present two of our coil designs recently developed for human head coil for 10.5T MRI scanners. They are the eight-shaped and the double-loop segmented coils. They both show homogeneous magnetic field distribution with good input impedance matching at the corresponding Larmor frequency. Homogeneous coil sensitivities are achieved.
  • Keywords
    "Magnetic resonance imaging","Magnetic fields","Radio frequency","Head","Magnetic heads","Radiofrequency identification"