Title :
Multi-focus tissue harmonic images obtained with parallel transmit beamforming by means of orthogonal frequency division multiplexing
Author :
Libertario Demi;Gabriele Giannini;Alessandro Ramalli;Piero Tortoli;Massimo Mischi
Author_Institution :
Lab. of Biomed. Diagnostics, Eindhoven Univ. of Technol., Eindhoven, Netherlands
Abstract :
Parallel beamforming in transmit (PBT) by means of orthogonal frequency division multiplexing (OFDM) was recently applied to increase the frame rate of ultrasound tissue harmonic imaging. OFDM-PBT improves the data acquisition rate by generating, for each transmission event, multiple beams that are allocated to a specific portion of the available transducer bandwidth. Improving the frame rate is however not the only possible application of an increased data acquisition rate. In this paper, OFDM-PBT is exploited to perform multi-focus tissue harmonic imaging for improving the penetration depth and signal to noise ratio (SNR) without affecting the frame rate. Tissue harmonic images of a tissue mimicking phantom were obtained with OFDM-PBT and standard B-mode imaging. Results present improved penetration depth and SNR for OFDM-PBT as compared to standard B-mode. We estimated an improvement of 6 dB of the average SNR, at the expenses of a reduction of the axial resolution (0.7 vs 1.1 mm).
Keywords :
"OFDM","Imaging","Standards","Signal to noise ratio","Harmonic analysis","Ultrasonic imaging","Image resolution"
Conference_Titel :
Ultrasonics Symposium (IUS), 2015 IEEE International
DOI :
10.1109/ULTSYM.2015.0501