• DocumentCode
    3692175
  • Title

    Fabrication and performance of a micro 50-MHz IVUS Transducer based on a 1-3 composite with geometric focusing

  • Author

    Xiaohua Jian; Zhile Han; Pengbo Liu; Zhangjian Li; Peiyang Li; Weiwei Shao; Yaoyao Cui

  • Author_Institution
    Suzhou Institute of Biomedical Engineering and Technology, CAS, China
  • fYear
    2015
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    In this paper a micro 50-MHz IVUS Transducer based on a 1-3 piezo-composite with geometric focusing was presented. The using 1-3 PZT composite material was developed in our laboratory with 12 um kerfs and 18 um pillars. The transducer was focused in the elevation direction by geometrically shaping the composite material using a PTEF ball with a 5 mm radius. After the silver conductive backing layer was cured, the transducer was divided into the size of 1.0 × 1.0 mm. The final fabricated transducer was fixed in a needle tip, and its diameter is 9F. According to the testing results with our homemade IVUS system, its lateral resolution is about 120 um, which effectively improved the lateral resolution of current IVUS system.
  • Keywords
    "Image resolution","Ultrasonic imaging","Imaging","Resonant frequency","Acoustics","Transducers","Ceramics"
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium (IUS), 2015 IEEE International
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2015.0551
  • Filename
    7329163