DocumentCode :
370
Title :
Combined Electro-Optical Imaging for the Time Evolution of White Thrombus Growth in Artificial Capillaries
Author :
Affanni, A. ; Specogna, Ruben ; Trevisan, F.
Author_Institution :
Dept. of Electr., Univ. of Udine, Udine, Italy
Volume :
62
Issue :
11
fYear :
2013
fDate :
Nov. 2013
Firstpage :
2954
Lastpage :
2959
Abstract :
We present a novel methodology to measure white thrombus volume growth in an artificial microchannel, where whole blood flows. We designed a sensor consisting of a poly-dimethyl-siloxane microchannel and parallel gold electrodes sputtered on the surface of a slide, where induced hemostasis takes place. A novel inversion methodology, based on optical and electrical impedance data simultaneously processed, allows reconstructing the thrombus volume. The advantage of the proposed methodology is to reconstruct the evolution of the thrombus volume as a function of time; this is not possible with the present state of the art optical imaging based on confocal microscopy, which provides the thrombus volume estimation only at the end of the process.
Keywords :
blood; blood vessels; electric impedance imaging; electro-optical effects; gold; haemodynamics; microchannel flow; artificial capillaries; artificial microchannel; blood flow; electrical impedance data; electrooptical imaging; hemostasis; parallel gold electrodes; poly-dimethyl-siloxane microchannel; time evolution; white thrombus growth; Electrical impedance spectroscopy; whole blood hemostasis;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.2013.2266012
Filename :
6542756
Link To Document :
بازگشت