DocumentCode :
3704479
Title :
Q-factor change of cavity by sample installation in resonance perturbation method
Author :
Taro Miura;Ken Tahara;Jun-ichi Sugiyama;Masahiro Horibe
Author_Institution :
The Mueller, JAPAN
fYear :
2015
Firstpage :
738
Lastpage :
741
Abstract :
In the measuring process of the cavity resonance perturbation method to evaluate permittivity, it has historically been supposed that Q-factor of the cavity itself is not affected by the installation of a sample since the method has been proposed. Recently, several experimental results to show the change of Q-factor by the installation of a sample were found. Measurements and discussions to confirm the change have been made and a method to obtain the intrinsic loss factor of the sample separated from Q-factor change of the cavity is proposed in this paper.
Keywords :
"Cavity resonators","Q-factor","Loss measurement","Tuners","Current density","Resonant frequency","Dielectric loss measurement"