DocumentCode :
3704486
Title :
Development of a 27.12 MHz CC-LCF intraluminal applicator for hyperthermia of the esophagus
Author :
J Crezee;HJ Harms;HJ Boluijt;J Sijbrands;MCCM Hulshof;HP Kok
Author_Institution :
Dept. Radiation Oncology, Academic Medical Center, Amsterdam, The Netherlands
fYear :
2015
Firstpage :
769
Lastpage :
772
Abstract :
Esophageal tumors can be treated with radiation combined with hyperthermia. This study investigated the design of a 27.12 MHz capacitively coupled intraluminal applicator for hyperthermia of the esophagus to be used in combination with external locoregional heating. It consists of four independent electrodes mounted on an inflatable balloon catheter, 10mm each, Measured SAR penetration depth is 4.5mm. Spatial steering is excellent and will prove useful in heterogeneous tumors. Combination with locoregional heating showed a strong increase in temperature rise. Next step is application in esophageal cancer.
Keywords :
"Applicators","Hyperthermia","Electrodes","Temperature measurement","Tumors","Heating","Temperature distribution"
Publisher :
ieee
Conference_Titel :
Microwave Conference (EuMC), 2015 European
Type :
conf
DOI :
10.1109/EuMC.2015.7345877