DocumentCode :
3706198
Title :
On how to extract breathing rate from PPG signal using wearable devices
Author :
A. Fusco;D. Locatelli;F. Onorati;G. C. Durelli;M. D. Santambrogio
Author_Institution :
DEIB, Politecnico di Milano, Milan, Italy
fYear :
2015
Firstpage :
1
Lastpage :
4
Abstract :
In this paper, an algorithm for breathing rate extraction from PPG signal is proposed. Two critical aspects have been endorsed during the implementation: i) good performances and ii) low computational complexity. The proposed solution is based on the Empirical Mode Decomposition (EMD) approach and it proves to be robust and accurate even in presence of noisy epochs. It has been validated on two distinct datasets: a)experimental data we have collected using wearables for physiological monitoring and b) recording sessions from PhysioBank MIMIC II Waveform Database. The presented results showed a mean absolute error of 0.0044 Hz, corresponding to 0.26 breaths per minute.
Keywords :
"Biomedical monitoring","Monitoring","Signal processing algorithms","Algorithm design and analysis","Databases","Correlation","Optical transmitters"
Publisher :
ieee
Conference_Titel :
Biomedical Circuits and Systems Conference (BioCAS), 2015 IEEE
Type :
conf
DOI :
10.1109/BioCAS.2015.7348369
Filename :
7348369
Link To Document :