DocumentCode
3706213
Title
An ECG dataset representing real-world signal characteristics for wearable computers
Author
Qingxue Zhang;Chakameh Zahed;Viswam Nathan;Drew A. Hall;Roozbeh Jafari
Author_Institution
University of Texas at Dallas
fYear
2015
Firstpage
1
Lastpage
4
Abstract
We present an ECG dataset collected in real-world scenarios for wearable devices that includes over 260 recordings of 90-210 seconds that provide guidance for designers to evaluate signal acquisition circuit and system solutions. Several variations on the signal acquisition path are demonstrated, including various sources of interference (baseline wander, motion artifacts, and power line interference), signal path variations (electrode type, coupling method, and common-mode rejection method), and electrode placements (wrist and chest). Based on detailed analysis of signal characteristics under different scenarios, analog front-end (AFE) design recommendations are proposed.
Keywords
"Electrocardiography","Electrodes","Couplings","Wrist","Interference","Biomedical monitoring","Dynamic range"
Publisher
ieee
Conference_Titel
Biomedical Circuits and Systems Conference (BioCAS), 2015 IEEE
Type
conf
DOI
10.1109/BioCAS.2015.7348384
Filename
7348384
Link To Document