DocumentCode :
3717623
Title :
Consistency of surface pulse and reciprocity calibration of piezoelectric AE sensors
Author :
Yu-ying Wang;Long-biao He;Hai-jiang Zhu;Ping Yang
Author_Institution :
Beijing University of Chemical Technology, 100029, China
fYear :
2015
Firstpage :
189
Lastpage :
192
Abstract :
The calibration of piezoelectric acoustic emission sensors mainly includes two methods, reciprocity method and surface pulse method. The former uses the principle of electro-acoustic reciprocity of piezoelectric sensors, and the latter needs a reference capacitance or laser system for displacement or velocity measurement. Prior studies have indicated that the two methods had some consistency, but some scholars still questioned the reciprocity calibration of piezoelectric acoustic emission sensors. In this paper the equivalent and real reciprocity coefficient were calculated and compared to test their consistency. Experimental system was established to test and verify the results. The test results showed that the reciprocity and surface pulse method had good agreement.
Keywords :
"Sensors","Calibration","Acoustic emission","Sensitivity","Surface impedance","Transmitters","Velocity measurement"
Publisher :
ieee
Conference_Titel :
Piezoelectricity, Acoustic Waves, and Device Applications (SPAWDA), 2015 Symposium on
Type :
conf
DOI :
10.1109/SPAWDA.2015.7364469
Filename :
7364469
Link To Document :