• DocumentCode
    3717629
  • Title

    Dual-focus scanning in volumetric HIFU ablation: Preliminary simulation study

  • Author

    Ke Li;Jing-feng Bai;Ya-zhu Chen;Xiang Ji

  • Author_Institution
    School of Biomedical Engineering, Shanghai Jiao Tong University, 200030, China
  • fYear
    2015
  • Firstpage
    216
  • Lastpage
    218
  • Abstract
    Volumetric HIFU ablation has been implemented in clinical practice to create large thermal lesion while reducing treatment durations. However, it still takes several hours to cover the whole target using volumetric ablation. Multiple focus intensity patterns have been proved beneficial to create larger lesions. The feasibility of dual-focus scanning in volumetric ablation has been evaluated. Simulations of dual-focus scanning and single-focus scanning were performed with the treatment cells of 4, 8, 12 and 16 mm in diameter. Comparison of both scanning approaches indicates that the dual-focus scanning saves treatment time by 20% and needs lower peak acoustic pressure, therefore the potential cavitation could be avoided. Although the maximum near-field temperatures are slightly higher than single-focus scanning, they are acceptable in HIFU treatment.
  • Keywords
    "Trajectory","Heating","Ultrasonic imaging","Temperature distribution","Piezoelectricity","Acoustic waves"
  • Publisher
    ieee
  • Conference_Titel
    Piezoelectricity, Acoustic Waves, and Device Applications (SPAWDA), 2015 Symposium on
  • Type

    conf

  • DOI
    10.1109/SPAWDA.2015.7364475
  • Filename
    7364475