DocumentCode :
3721954
Title :
Integration of fractal biosensor in a digital microfluidic platform
Author :
Yousof Mashraei;Shilpa Sivashankar;Ulrich Buttner;Khaled Nabli Salama
Author_Institution :
Computer, Electrical And Mathematical Sciences and Engineering (CEMSE) King Abdullah University of Science and Technology (KAUST) Thwual 23955-6900, Kingdom of Saudi Arabia
fYear :
2015
Firstpage :
1
Lastpage :
4
Abstract :
Fractal capacitive electrodes have been successfully integrated into a digital microfluidic open-platform. These electrodes perform actuation and withstand voltages up to 300V without insulation-layer breakdown. They were used to quantify the concentration levels of C-reactive protein (CRP) to determine the risk of cardiovascular disease. The capacitance increased sevenfold and stabilized in less than 5 minutes. The sensor shows a decreasing trend of capacitance readouts with the increase of concentrations. The same immunoassay was tested with untreated electrodes and showed no significant response, which suggests that immobilization was necessary. This configuration allows the electrodes to be used as biosensors.
Keywords :
"Microfluidics","Capacitance","Biosensors","Electrodes","Metals","Resists","Surface treatment"
Publisher :
ieee
Conference_Titel :
SENSORS, 2015 IEEE
Type :
conf
DOI :
10.1109/ICSENS.2015.7370497
Filename :
7370497
Link To Document :