• DocumentCode
    3722090
  • Title

    Birdcage type NMR receiver coil sensor with integrated detuning circuit for 3T MRI system

  • Author

    Sheikh Faisal Ahmad;Young Cheol Kim;Ick Chang Choi;Hyun Deok Kim

  • Author_Institution
    Department of Electronics Engineering, Kyungpook National University, Daegu, South Korea
  • fYear
    2015
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    A novel birdcage type Radio Frequency (RF) receiver coil sensor with passive type detuning circuits integrated on the end-rings of the coil sensor has been implemented. A simulation model of the RF receiver coil sensor with the passive type detuning circuits has also been proposed. The simulation model completely explains the working principle of the RF receiver coil sensor with the passive type detuning circuits. The RF coil sensor was designed to perform the Nuclear Magnetic Resonance (NMR) imaging of small animal at 3T Magnetic Resonance Imaging (MRI) system. The integrated detuning circuits of the RF coil sensor eliminate the noises induced by the mutual inductance between receiving and transmitting coils. We have confirmed the performances of the RF coil sensor through experiments, which agreed well with those of the simulations.
  • Keywords
    "Radio frequency","Integrated circuit modeling","Receivers","Nuclear magnetic resonance","Magnetic resonance imaging","Capacitors","Switching circuits"
  • Publisher
    ieee
  • Conference_Titel
    SENSORS, 2015 IEEE
  • Type

    conf

  • DOI
    10.1109/ICSENS.2015.7370638
  • Filename
    7370638