DocumentCode :
3724230
Title :
Design and testing of a gentle pump with rotating magnetic field for fluids with stress-sensitive microparticles
Author :
Mohammad Kilani;Hussam Khasawneh;Abdallah Awidi Abbadi
Author_Institution :
The University of Jordan, Department of Mechatronics, Amman 11942, Jordan
fYear :
2015
Firstpage :
1
Lastpage :
4
Abstract :
This paper presents the design and testing of a gentle pump prototype that is operated by a rotating magnetic field. The pump targets applications in which the pumped fluid contains stress sensitive microparticles which may be damaged when subjected to the high stress levels prevalent in conventional pumps. The prototype was tested in-vitro using blood from human volunteers, and according to published international protocols for the assessment of the hemolytic properties of continuous flow blood pumps used in extracorporeal or implantable circulatory assist devices. The normalized index of hemolysis (NIH) was found to be about 0.007, which is comparable with that obtained from top ranking blood pumps available in the market today.
Keywords :
"Blood","Pistons","DC motors","Prototypes","Pumps","Magnetic fields","Plasmas"
Publisher :
ieee
Conference_Titel :
Mechatronics and its Applications (ISMA), 2015 10th International Symposium on
Type :
conf
DOI :
10.1109/ISMA.2015.7373498
Filename :
7373498
Link To Document :