• DocumentCode
    3731493
  • Title

    Current-Biased Kinetic Inductance Detector for Neutrons

  • Author

    Hiroaki Shishido;Shigeyuki Miyajima;Hiroyuki Yamaguchi;Hirotaka Nakayama;Yoshito Narukami;Akira Fujimaki;Mutsuo Hidaka;Ken´ichi Oikawa;Masahide Harada;Takayuki Oku;Masatoshi Arai;Takekazu Ishida

  • Author_Institution
    Dept. of Phys. &
  • fYear
    2015
  • fDate
    7/1/2015 12:00:00 AM
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    We succeeded in measuring a clear signal from a single neutron by using a solid state superconducting current biased kinetic inductance detector, which consists of a superconducting Nb meanderline and an enriched 10B neutron conversion layer. Our neutron detector is able to operate at temperatures much lower than Tc. A full width at half maximum of signal remains at a few tens ns, ensuring fast operation as a neutron detector. A time-dependent histogram for the pulsed neutron intensity is in good agreement with the prediction from a Monte-Carlo simulation. We convincingly conclude that our detector was successful in detecting pulsed neutrons produced from the beam line BL10 at J-PARC.
  • Keywords
    "Neutrons","Detectors","Niobium","Temperature control","Superconducting epitaxial layers","Inductance","Temperature measurement"
  • Publisher
    ieee
  • Conference_Titel
    Superconductive Electronics Conference (ISEC), 2015 15th International
  • Type

    conf

  • DOI
    10.1109/ISEC.2015.7383436
  • Filename
    7383436