• DocumentCode
    3731541
  • Title

    Thermal-Converter Validation of Differential Sampling Measurement Based on AC-Programmable Josephson Voltage Standard System

  • Author

    Y. Amagai;M. Maruyama;H. Yamamori;C. Urano;S.-F. Chen;H. Fujiki;N.-H. Kaneko

  • Author_Institution
    Nat. Inst. of Adv. Ind. Sci. &
  • fYear
    2015
  • fDate
    7/1/2015 12:00:00 AM
  • Firstpage
    1
  • Lastpage
    3
  • Abstract
    This paper describes a differential sampling measurement system using AC-programmable Josephson voltage standard (AC-PJVS) system up to 10 V in the frequency range down to 1 Hz. Such system combines the versatility and the accuracy and stability of the Josephson voltage standard. The validity of this method is confirmed with the AC-DC difference measurements of a calibrated thermal converter, and with an electro-thermal simulation at the root-mean square amplitudes of 3 V. The uncertainty of our differential sampling measurement system is estimated to be 1.3 μV/V for waveform frequencies of 62.5 Hz and root-mean square amplitude of 10 V.
  • Keywords
    "Voltage measurement","Frequency measurement","Uncertainty","Standards","Semiconductor device measurement","Voltmeters","Measurement uncertainty"