Title :
Study on measurement method of luminous intensity based on the third generation low light level image intensifier
Author :
Wang Rui;Bai Xiao-feng;Wang Rui;Bai Xiao-feng
Author_Institution :
Science and Technology on Low-Light-Level Night Vision Laboratory, Xi´an 710065, China
fDate :
7/1/2015 12:00:00 AM
Abstract :
This paper analyzed the technical characteristics of GaAs photocathode electrode with high sensitivity. Through measuring the output current of micro channel plate, the relation between incident light intensity and output current is summarized. Besides, we proposed the light intensity measurement method for the third generation low light level (LLL) image intensifier. By comparing our method with the method of measuring the intensity of the photoelectric multiplier tube, our method has the advantages of high sensitivity, fast response, independent of temperature, and measurement simplicity. The results indicate implicit linearity between micro channel plate and intensity of light under 10-4~10-1lx environment. It verifies the validness of our method.
Keywords :
"Current measurement","Lighting","Image intensifiers","Gallium arsenide","Cathodes","Electron tubes","Voltage measurement"
Conference_Titel :
Optoelectronics and Microelectronics (ICOM), 2015 International Conference on
DOI :
10.1109/ICoOM.2015.7398864