• DocumentCode
    3742404
  • Title

    The magnetic induction tomography measurement system based on Helmholtz coil

  • Author

    Haijun Luo;Xin Jiang

  • Author_Institution
    College of Physics and Electronic engineering, Chongqing Normal University, Chongqing, China, 401331
  • fYear
    2015
  • Firstpage
    29
  • Lastpage
    33
  • Abstract
    Magnetic induction tomography (MIT) is a contactless measurement of biological tissue conductivity imaging technique. The spiral coil is widely used, which leads to the nonuniformity of sensitivity distribution of the detection signal. This paper proposed the Helmholtz coil which can generate a set of uniform excitation fields, and analyzed the uniformity of excitation field. In MIT system, the exciting magnetic field is generated through sine-wave current, with the frequency of 10MHz and exciting current of 20mA. The experimental physics model simulated cerebral hemorrhage is produced by using the distilled water, the agar powder and NaCl. The result experiments show that this system can test the changes of the conductivity of partial tissues, and reflect the position of the foreign matter, but the image is enlarged compared with the actual one.
  • Keywords
    "Magnetic fields","Conductivity","Magnetic field measurement","Biomedical measurement","Magnetic resonance imaging","Current measurement"
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Engineering and Informatics (BMEI), 2015 8th International Conference on
  • Type

    conf

  • DOI
    10.1109/BMEI.2015.7401468
  • Filename
    7401468