• DocumentCode
    3745154
  • Title

    Tactile Imaging System for inclusion size and stiffness characterization

  • Author

    Vira Oleksyuk;Firdous Saleheen; Yi Chen;Chang-Hee Won

  • Author_Institution
    Department of Electrical and Computer Engineering, Temple University, Philadelphia, PA 19122, USA
  • fYear
    2015
  • Firstpage
    1
  • Lastpage
    6
  • Abstract
    We developed Tactile Imaging System (TIS), which measures mechanical properties of tissue inclusions. The functional components of TIS include: a monochrome camera, a soft and transparent silicone probe, and an LED illumination circuit. The deformation of the soft TIS probe is used to evaluate size and compliance of inclusions, which is the inverse of elastic modulus. TIS algorithm development and testing were performed using a fabricated silicone tissue phantom with soft inclusions. We completed validation experiments for size and compliance estimation with TIS using the phantom. The results from these experiments suggest that TIS can be used for characterization of tissue inclusions.
  • Keywords
    "Light emitting diodes","Force","Image resolution","Size measurement","Lighting","Cameras","Switches"
  • Publisher
    ieee
  • Conference_Titel
    Signal Processing in Medicine and Biology Symposium (SPMB), 2015 IEEE
  • Type

    conf

  • DOI
    10.1109/SPMB.2015.7405472
  • Filename
    7405472