• DocumentCode
    3746482
  • Title

    Evaluation of Modulation Transfer Function on the optical tomography imaging system

  • Author

    Yicheng Wang;Baosheng Li

  • Author_Institution
    School of Instrument Science & Opto-electric Engineering, Hefei University of Technology, Hefei, Anhui, China
  • fYear
    2015
  • Firstpage
    717
  • Lastpage
    721
  • Abstract
    The optical tomography system based on the single pixel detector produces the high resolution image with optical scanning. This paper introduces a method to evaluate the quality of the optical tomography system with MTF (Modulation Transfer Function). At first use the calibration board that are digital images for software simulation, and then calculate the MTF values on the different space frequency in order to analyze the change of MTF and the reasons of image degradation. Followed by using the integrating sphere light source and film calibration boards in experiments, it tests the quality of the tomography imaging system and the validation of imaging properties. The system produces the images with Filter Back Projection (FBP). It analyzes image diffusion situation when the target image is the round hole. The results suggest that the effect of algorithm on the reconstructed image for the optical tomography system cannot be ignored.
  • Keywords
    "Optical filters","Optical imaging","Tomography","Image reconstruction","Optical films","Optical reflection"
  • Publisher
    ieee
  • Conference_Titel
    Image and Signal Processing (CISP), 2015 8th International Congress on
  • Type

    conf

  • DOI
    10.1109/CISP.2015.7407971
  • Filename
    7407971