• DocumentCode
    374771
  • Title

    The impact of noisy attenuation maps and patient motion on human-observer performance at Ga-67 lesion detection in SPECT

  • Author

    Wells, R.G. ; Gifford, H.C. ; Pretorius, P.H. ; Farncombe, T. ; King, M.A.

  • Author_Institution
    Lawson Health Res. Inst., London, Ont., Canada
  • Volume
    2
  • fYear
    2000
  • fDate
    2000
  • Abstract
    The authors have demonstrated an improvement due to attenuation correction (AC) at the task of lesion detection in thoracic SPECT images. However, increased noise in the transmission data due to aging sources or very large patients, and misregistration of the emission and transmission maps, can reduce the benefits of AC and may result in a loss of lesion detectability. The authors investigated the impact of noise in and misregistration of transmission data, on the detection of Ga-67 thoracic lesions. Human-observer LROC methodology was used to assess performance. Both emission and transmission data were simulated using the MCAT computer phantom. Images were reconstructed using OSEM incorporating AC and detector resolution compensation. Clinical count levels were used in the emission data. The transmission-data noise levels ranged from zero (noise-free) to 32 times measured clinical levels. Transaxial misregistrations of 0.32, 0.63, and 1.27 cm between emission and transmission data were also examined. Results indicate that a 20-fold increase in the noise was required to eliminate the benefit afforded by AC but that smaller increases in noise could be detrimental, especially for low-contrast lesions. Misregistration errors are also a concern as even small errors here greatly reduce the performance gains of AC
  • Keywords
    biomechanics; gamma-ray absorption; image registration; medical image processing; single photon emission computed tomography; visual perception; Ga; Ga-67 lesion detection; MCAT computer phantom; aging sources; clinical count levels; detector resolution compensation; human-observer performance; medical diagnostic imaging; misregistration; noisy attenuation maps; nuclear medicine; patient motion; performance gains; thoracic SPECT images; transaxial misregistrations; very large patient; Aging; Attenuation; Computational modeling; Computer simulation; Image reconstruction; Imaging phantoms; Lesions; Noise level; Noise reduction; Propagation losses;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2000 IEEE
  • Conference_Location
    Lyon
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-6503-8
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2000.949981
  • Filename
    949981