Image properties of list mode likelihood reconstruction for a rectangular positron emission mammograph with DOI measurements
Author
Qi, Jinyi ; Huesman, Ronald H.
Author_Institution
Lawrence Berkeley Nat. Lab., CA, USA
Volume
2
fYear
2000
fDate
2000
Abstract
A positron emission mammography scanner is under development at our Laboratory. The tomograph has a rectangular geometry consisting of four banks of detector modules. For each detector, the system can measure the depth of interaction information inside the crystal. The rectangular geometry leads to irregular radial and angular sampling and spatially variant sensitivity that are different from conventional PET systems. Therefore, it is of importance to study the image properties of the reconstructions. We adapted the theoretical analysis that we had developed for conventional PET systems to the list mode likelihood reconstruction for this tomograph. The local impulse response and covariance of the reconstruction can be easily computed using FFT. These theoretical results are also used with computer observer models to compute the signal-to-noise ratio for lesion detection. The analysis reveals the spatially variant resolution and noise properties of the list mode likelihood reconstruction. The theoretical predictions are in good agreement with Monte Carlo results
Keywords
Monte Carlo methods; image reconstruction; image resolution; mammography; medical image processing; positron emission tomography; tumours; DOI measurements; FFT; Monte Carlo results; computer observer models; covariance; detector modules; image properties; irregular angular sampling; irregular radial sampling; lesion detection; list mode likelihood reconstruction; local impulse response; noise properties; positron emission mammography scanner; rectangular geometry; rectangular positron emission mammograph; signal-to-noise ratio; spatially variant resolution; spatially variant sensitivity; Detectors; Geometry; Image reconstruction; Image sampling; Laboratories; Lesions; Mammography; Positron emission tomography; Radioactive decay; Signal to noise ratio;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record, 2000 IEEE
Conference_Location
Lyon
ISSN
1082-3654
Print_ISBN
0-7803-6503-8
Type
conf
DOI
10.1109/NSSMIC.2000.950047
Filename
950047
Link To Document