• DocumentCode
    3749033
  • Title

    Adaptive step size LMS for ECG artefact reduction during MRI

  • Author

    A Guillou;S M?n?tr?;G Petitmangin;J Felblinger;L Bonnemains

  • Author_Institution
    Schiller, Wissembourg, France
  • fYear
    2015
  • Firstpage
    761
  • Lastpage
    764
  • Abstract
    During cardiac MRI, fast switching gradients cause artifacts on the electrocardiogram (ECG), disturbing both triggering and patient monitoring. To cancel this noise, the Least Mean Squares (LMS) algorithm is a simple and efficient method. LMS uses one main parameter, its step size, which influences the quality of artifact reduction. We propose a method using the MR gradient variance to choose this parameter accurately using information about the sequence played by the MR scanner. The proposed method achieved systematically better results than the standard LMS with a 1.5T ECG database.
  • Keywords
    "Magnetic resonance imaging","Databases","Lead","Heart","Sociology","Statistics"
  • Publisher
    ieee
  • Conference_Titel
    Computing in Cardiology Conference (CinC), 2015
  • ISSN
    2325-8861
  • Print_ISBN
    978-1-5090-0685-4
  • Electronic_ISBN
    2325-887X
  • Type

    conf

  • DOI
    10.1109/CIC.2015.7411022
  • Filename
    7411022