• DocumentCode
    3755732
  • Title

    Dithered multi-pulsing and non-parametric statistical inference algorithm for time-of-flight mass spectrometry

  • Author

    George S. Moore;Michael Manlove;August Hidalgo

  • Author_Institution
    Keysight Laboratories, Keysight Technologies, Santa Clara, CA 95051
  • fYear
    2015
  • Firstpage
    531
  • Lastpage
    535
  • Abstract
    Time-of-flight mass spectrometers have classically been limited in the maximum ion pulser firing rate by the flight time of the slowest analyte ion present. This paper describes a new randomly dithered ion pulser firing scheme that operates in conjunction with a non- parametric iterated statistical inference algorithm, inspired by the Belief Propagation algorithm, to resolve the multiple time-of-flight aliases that result for firing rates above this classical limit. This algorithm readily exploits available a priori information and several acceleration techniques are applicable. Simulations of this method with a 10x firing rate increases show excellent fidelity with classically measured spectrums.
  • Keywords
    Decision support systems
  • Publisher
    ieee
  • Conference_Titel
    Signals, Systems and Computers, 2015 49th Asilomar Conference on
  • Electronic_ISBN
    1058-6393
  • Type

    conf

  • DOI
    10.1109/ACSSC.2015.7421185
  • Filename
    7421185