DocumentCode :
3759612
Title :
Calibration and data standardization of a prototype bench-top preclinical CT
Author :
Richard Taschereau;Nam T. Vu;Arion F. Chatziioannou
Author_Institution :
Crump Institute for Molecular Imaging, UCLA School of Medicine, Department of Molecular and Medical Pharmacology, Los Angeles, CA 90095 USA
fYear :
2014
Firstpage :
1
Lastpage :
2
Abstract :
This paper reports on the geometric calibration procedure of a bench-top, high-speed, and continuous-rotation preclinical computed tomography (CT) scanner that has been developed at our institution. A full 13-parameter model of the image formation process is described. Once calibrated, the model is used to modify projection data sets - the standardization operation - so that they appear to come from an ideal system. This technique has the advantage that standardized scans can be fed to any reconstruction algorithm since no awareness of the actual system geometry is required from the reconstruction program.
Keywords :
"Detectors","Calibration","Computed tomography","Image reconstruction","Phantoms","Standardization","Geometry"
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2014 IEEE
Type :
conf
DOI :
10.1109/NSSMIC.2014.7430845
Filename :
7430845
Link To Document :