Title :
Characterization of a YAG:Ce calorimeter with high-energy proton beam
Author :
V. Sipala;N. Randazzo;S. Aiello;M. Bruzzi;M. Bucciolini;M. Carpinelli;G. A. P. Cirrone;C. Civinini;G. Cuttone;E. Leonora;D. Lo Presti;S. Pallotta;M. Scaringella;C. Stancampiano;C. Talamonti;E. Vanzi
Author_Institution :
INFN of Catania, Italy
Abstract :
A YAG:Ce calorimeter with a 6×6 cm2 active area, long 10 cm has been manufactured in order to stop up to 200 MeV protons and to sustain 1MHz particle rate. It consists of four optically separated crystals assembled in the same housing. The YAG:Ce scintillator is promising for charged particle detection applications where high-count rate, good energy resolution and compact photodiode readout, not influenced by magnetic fields, are of importance. The YAG:Ce calorimeter has been characterized with proton beams up to 175MeV a The Svedberg Laboratory (TSL), Uppsala Sweden. Results obtained are shown and discussed.
Keywords :
"Crystals","Protons","Particle beams","Photodiodes","Energy resolution","Computed tomography","Silicon"
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2014 IEEE
DOI :
10.1109/NSSMIC.2014.7431200