• DocumentCode
    3762009
  • Title

    Simulation of DAC-based truncated sine excitation pulse generator

  • Author

    A. Aliakbari;Yousef Mohammadi Yeganeh;S. Safari

  • Author_Institution
    Faculty of Electrical Engineering, K. N. Toosi University of Technology, Tehran, Iran
  • fYear
    2015
  • Firstpage
    689
  • Lastpage
    693
  • Abstract
    A signal with high quality rectangular frequency spectrum is generated for MRI medical imaging applications. The center frequency of the signal is 4MHz and the bandwidth is 8KHz. This signal is generated by feeding Digital to Analog Converter (DAC) with repetitive truncated and modulated sinc function. Resolution and update rate of the required DAC is chosen theoretically and by simulations in MATLAB. Further simulations are carried out to optimize the truncating, which resulted in 800Hz transition bandwidth, -32.27dB pass-band ripple, -37.92dB stop-band ripple and 6.977MHz image signal gap by means of feeding a 15MSPS 14-bit DAC with 75,000 samples.
  • Keywords
    "Decision support systems","Quantization (signal)","Time-domain analysis"
  • Publisher
    ieee
  • Conference_Titel
    Knowledge-Based Engineering and Innovation (KBEI), 2015 2nd International Conference on
  • Type

    conf

  • DOI
    10.1109/KBEI.2015.7436127
  • Filename
    7436127