• DocumentCode
    3766232
  • Title

    Challenges in the localization of a UHF radio pill in the gastro-intestinal tract

  • Author

    Camille Aubry;Hugo G. Espinosa;David V. Thiel;Esmaeil S. Nadimi

  • Author_Institution
    Biomedical Engineering, Faculty of Medicine Nancy. University of Lorraine, Nancy, France
  • fYear
    2015
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    Visual and chemical investigations in the gastro-intestinal tract using injected "pill" sized sensors are challenging. The throughput takes between 24 and 48 hours and a precise location accuracy to less than 5 cm is necessary for keyhole surgery. Model and computational experiments using a 433 MHz subsurface transmitter in a saline (conductivity = 0.2 S/m) tank show that location accuracy is compromised by non-conducting boundaries between internal organs connected by non-conducting interstitial tissue. The internal transmitter profile changes significantly by the presence of insulating boundaries modelled by non-conducting boundary in an otherwise uniform conducting medium. A worn sensor array in the front of the abdomen can define the location of a GI pill only if there is significant information about the internal organ configuration.
  • Keywords
    "Conductivity","Radio transmitters","Electric fields","Containers","Receivers","Biological systems"
  • Publisher
    ieee
  • Conference_Titel
    Antennas and Propagation (ISAP), 2015 International Symposium
  • Type

    conf

  • Filename
    7447300