DocumentCode :
37715
Title :
Characterization of Force and Torque Interactions During a Simulated Transgastric Appendectomy Procedure
Author :
Dargar, Saurabh ; Brino, Cecilia ; Matthes, Kai ; Sankaranarayanan, Ganesh ; De, Suvranu
Volume :
62
Issue :
3
fYear :
2015
fDate :
Mar-15
Firstpage :
890
Lastpage :
899
Abstract :
We have developed an instrumented endoscope grip handle equipped with a six-axis load cell and measured forces and torques during a simulated transgastric natural orifice translumenal endoscopic surgery appendectomy procedure performed in an EASIE-R ex vivo simulator. The data were collected from ten participating surgeons of varying degrees of expertise which was analyzed to compute a set of six force and torque parameters for each coordinate axis for each of the nine tasks of the appendectomy procedure. The mean push/pull force was found to be 3.64 N (σ = 3.54 N) in the push direction and the mean torque was 3.3 N · mm (σ = 38.6 N · mm) in the counterclockwise direction about the push/pull axis. Most interestingly, the force and torque data about the nondominant x and z axes showed a statistically significant difference (p <; 0.05) between the expert and novice groups for five of the nine tasks. These data may be useful in developing surgical platforms especially new haptic devices and simulation systems for emerging natural orifice procedures.
Keywords :
endoscopes; haptic interfaces; surgery; endoscope grip handle; force interaction; haptic device; natural orifice procedure; natural orifice translumenal endoscopic surgery; simulated transgastric appendectomy procedure; six-axis load cell; torque interaction; Biomedical measurement; Endoscopes; Force; Force measurement; Haptic interfaces; Surgery; Torque; Haptic interface; human factors; surgery; virtual reality;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/TBME.2014.2369956
Filename :
6954427
Link To Document :
بازگشت