DocumentCode :
3772809
Title :
Study of Fricke-gel dosimeter calibration for attaining precise measurements of the absorbed dose
Author :
G. M. Liosi;G. Gambarini;E. Artuso;S. Benedini;E. Macerata;F. Giacobbo;M. Gargano;N. Ludwig;M. Carrara;E. Pignoli;M. Mariani
Author_Institution :
Department of Energy, Nuclear Engineering Division, Politecnico di Milano, Milan, Italy
fYear :
2015
fDate :
4/1/2015 12:00:00 AM
Firstpage :
1
Lastpage :
5
Abstract :
A calibration method has been studied for attaining, with good precision, absolute measurements of the spatial distribution of the absorbed dose by means of the Fricke gelatin xylenol orange dosimetric system. With this aim, the dose response to subsequent irradiations was analyzed. In fact, the studied modality is based on a pre-irradiation of each single dosimeter in a uniform field with a known dose, in order to extrapolate a calibration image for a subsequent non-uniform irradiation with an unknown dose to be measured.
Keywords :
"Radiation effects","Calibration","Optical imaging","Optical variables measurement","Biomedical optical imaging","Absorption","Physics"