DocumentCode :
3772877
Title :
Thomson scattering diagnostic data acquisition systems for modern fusion systems
Author :
S. V. Ivanenko;A. D. Khilchenko;V. K. Ovchar;P. V. Zubarev;A. N. Kvashnin;E. A. Puryga;A. A. Ivanova;A. I. Kotelnikov
Author_Institution :
Budker Institute of Nuclear Physics and Novosibirsk State University, Novosibirsk, Russia
fYear :
2015
fDate :
4/1/2015 12:00:00 AM
Firstpage :
1
Lastpage :
4
Abstract :
Uniquely designed complex data acquisition system for Thomson scattering diagnostic was developed. It allows recording short duration (3-5 ns) scattered pulses with 2GHz sampling rate and 10-bit total resolution in oscilloscope mode. The system consists up to 48 photo detector modules with 0200MHz bandwidth, 1-48 simultaneously sampling ADC modules and synchronization subsystem. The photo detector modules are based on avalanche photodiodes (APD) and ultra-low noise transimpedance amplifiers. ADC modules include fast analog to digital convertors and digital units based on the FPGA (Field-Programmable Gate Array) for data processing and storage. The synchronization subsystem is used to form triggering pulses and to organize the simultaneously mode of ADC modules operation.
Keywords :
"Photodetectors","Data acquisition","Synchronization","Scattering","Clocks","Detectors","Field programmable gate arrays"
Publisher :
ieee
Conference_Titel :
Advancements in Nuclear Instrumentation Measurement Methods and their Applications (ANIMMA), 2015 4th International Conference on
Type :
conf
DOI :
10.1109/ANIMMA.2015.7465649