Title :
Plethysmography measurements using short current pulses with low-duty cycle
Author :
A. Santic;T. Stritof;V. Bilas
Author_Institution :
Fac. of Electr. Eng. & Comput., Zagreb Univ., Croatia
Abstract :
Pulse plethysmography with very low-duty cycle offers much lower current load of examined bio-tissue in comparison to sine-wave impedance plethysmography. This enables application of strong current pulses with the same bio-tissue loading effect as for impedance plethysmography, resulting in larger signal and better signal to noise ratio. Determination of an equivalent electrical circuit of electrodes connected with bio-tissue is possible by fitting the measured waveshape with a resistor and capacitor connected in series, or better with three or four elements. A simplified block diagram of the pulse plethysmograph is given as well. Blood pressure wave-shape obtained on arteries and volume changes during respiration are measured and shown.
Keywords :
"Pulse measurements","Plethysmography","Current measurement","Impedance","Signal to noise ratio","Circuits","Electrodes","Electric variables measurement","Resistors","Capacitors"
Conference_Titel :
Engineering in Medicine and Biology Society, 1998. Proceedings of the 20th Annual International Conference of the IEEE
Print_ISBN :
0-7803-5164-9
DOI :
10.1109/IEMBS.1998.746965