DocumentCode :
3777954
Title :
The realization of spectral total radiance flux scale and related calibration service at NIM
Author :
Zhao Weiqiang; Liu Hui; Liu Jian
Author_Institution :
National Institute of Metrology, Beijing, China 100029
Volume :
1
fYear :
2015
fDate :
7/1/2015 12:00:00 AM
Firstpage :
285
Lastpage :
289
Abstract :
National Institute of Metrology China (NIM) has studied and developed the facilities for spectral total radiant flux (STRF) measurement and provides related calibration services. The array-based goniospectroradiometers and an integrating sphere system have been setup. The former is used to realize the scale of the STRF, and the latter is used to implement everyday calibration service due to its fast speed. The STRF standard lamps are also developed. For the goniospectroradiometer, the measurement result has an uncertainty of 1.8% ~ 3.6% (k=2) within the spectral range of 380 nm to 780 nm. For the integrating sphere system, the measurement result has an uncertainty of 2.0% ~ 3.8% (k=2) within the spectral range of 380 nm to 780 nm.
Keywords :
"Standards","Light emitting diodes","Spectroradiometers","Calibration","Measurement uncertainty","Uncertainty","Light sources"
Publisher :
ieee
Conference_Titel :
Electronic Measurement & Instruments (ICEMI), 2015 12th IEEE International Conference on
Type :
conf
DOI :
10.1109/ICEMI.2015.7494174
Filename :
7494174
Link To Document :
بازگشت