• DocumentCode
    3778072
  • Title

    Operation point selection for thin-window large areas flow gas proportional counter

  • Author

    Guo Huiping; Lv Wenhui; Lv Ning; Zhao Kuo

  • Author_Institution
    Xi´an Research Insitute of Hi- Tech, 710025, China
  • Volume
    2
  • fYear
    2015
  • fDate
    7/1/2015 12:00:00 AM
  • Firstpage
    622
  • Lastpage
    626
  • Abstract
    In order to measure the surface contamination level effectively, the large area surface contamination measuring instrument is widely used. This research designs a thin-window flow gas proportional counter with 1000cm2 detecting area. When to achieve simultaneous measurement of α and β rays in using pulse-height discrimination technology, the selection of high voltage operation point for this detector can´t follow the traditional method. As the voltage increases, β pulse-height increases, string to α channel, making the result more inaccuracy. Based on experiments, considered the limitation of the interference ratio in α+β count aerodromes, the operation point of detector was located near the head of plateau. Experimental results proved that the selection methods of the working point meet the regulation of national standard.
  • Keywords
    "Radiation detectors","Pollution measurement","Atmospheric measurements","Particle measurements","Instruments","Surface contamination"
  • Publisher
    ieee
  • Conference_Titel
    Electronic Measurement & Instruments (ICEMI), 2015 12th IEEE International Conference on
  • Type

    conf

  • DOI
    10.1109/ICEMI.2015.7494296
  • Filename
    7494296