• DocumentCode
    3780000
  • Title

    A variable active cold noise source for calibration of medical microwave radiometer

  • Author

    Vidyalakshmi.M.R.;Kavitha Arunachalam

  • Author_Institution
    Dept. of Engineering Design, Indian Institute of Technology Madras, Chennai, India
  • fYear
    2015
  • Firstpage
    1
  • Lastpage
    2
  • Abstract
    An active cold noise source with low return loss and stable operating temperature of 288 K is presented for a microwave radiometer designed for medical application (288 K to 310 K). The noise source is based on Si-Ge Heterojunction Bipolar Transistor (HBT) and is proposed for calibration of our L-band radiometer designed for detection of thermal anomalies in the human body. The noise source is centered at 1.3 GHz with 190 MHz bandwidth for return loss less than -20 dB, 3.01 dB noise figure and variable noise temperature source for the low temperature reference (288 K - 298 K).
  • Keywords
    "Microwave radiometry","Temperature measurement","Microwave imaging","Loss measurement","Temperature sensors","Biomedical imaging","Microwave theory and techniques"
  • Publisher
    ieee
  • Conference_Titel
    Applied Electromagnetics Conference (AEMC), 2015 IEEE
  • Type

    conf

  • DOI
    10.1109/AEMC.2015.7509257
  • Filename
    7509257