• DocumentCode
    3780514
  • Title

    Research on the electromagnetic protection of high voltage cables in high speed EMU of power frequency magnetic field

  • Author

    Linsong Feng; Yinghong Wen; Jinbao Zhang; Xiaodong He

  • Author_Institution
    Beijing Jiaotong University Electromagnetic Compatibility Laboratory, 100044, China
  • fYear
    2015
  • Firstpage
    450
  • Lastpage
    453
  • Abstract
    Aiming at protect passengers with portable pacemaker, it is of vital importance to meet the standard in the design of EMU. In this paper, based on the theory of high voltage cable and electromagnetic protection, the model of high voltage cable passing through the compartment is established to simulate the magnetic flux density of the environment around the cable. In view of the actual situation, using sine, harmonic and inrush current as excitation source of high voltage cable under power frequency magnetic field. The results show that, under these conditions the magnetic field of the high voltage cable cannot affect the normal operation of pacemaker and the length of wire which is without shielding layer should be reduced as much as possible.
  • Keywords
    "Magnetic noise","Magnetic shielding","Cable shielding","Frequency measurement","Magnetic circuits","Analytical models"
  • Publisher
    ieee
  • Conference_Titel
    Microwave, Antenna, Propagation, and EMC Technologies (MAPE), 2015 IEEE 6th International Symposium on
  • Type

    conf

  • DOI
    10.1109/MAPE.2015.7510355
  • Filename
    7510355