Title :
A quadrature bridge for QHE-based calibration of capacitance standards
Author :
J. Bohacek;J. Holub;J. Roztocil;J. Vedral
Author_Institution :
Fac. of Electr. Eng., Czechoslovak Acad. of Sci., Prague, Czech Republic
Abstract :
A method is described which utilizes a 12906.4 /spl Omega/ resistor as a reference standard for capacitance calibrations. DC value of this resistor being obtained by comparing it against a quantum Hall effect device, its frequency dependence is measured by 1:1 comparison against a resistor having calculable frequency performance. Subsequently, the reference resistor and a 10 nF capacitor are linked by means of a quadrature bridge based on a commercial two-phase generator. The 10 nF capacitor, in turn, serves as a reference standard for calibration of capacitors of lower values by means of a four terminal-pair transformer bridge. By means of the quadrature bridge, capacitance can be measured with a standard uncertainty of less than 3 parts in 10/sup 5/ in the frequency range from 400 Hz to 5 kHz at the 10 nF level.
Keywords :
"Bridge circuits","Calibration","Resistors","Capacitors","Quantum capacitance","Frequency measurement","Hall effect devices","Frequency dependence","Capacitance measurement","Measurement standards"
Conference_Titel :
Instrumentation and Measurement Technology Conference, 1999. IMTC/99. Proceedings of the 16th IEEE
Print_ISBN :
0-7803-5276-9
DOI :
10.1109/IMTC.1999.777015