DocumentCode
3784720
Title
DVM input and leakage resistance analysis in the DVM-based resistance measurement systems
Author
R. Malaric;J. Butorac;I. Lenicek
Author_Institution
Dept. of Electr. Eng. Fundamentals & Meas., Univ. of Zagreb, Croatia
Volume
51
Issue
6
fYear
2002
Firstpage
1323
Lastpage
1327
Abstract
The digit voltmeter (DVM) input and leakage resistances have been analyzed in the DVM-based resistance measurement system using one or two DVMs. The results are given both theoretically and experimentally. Shunting errors due to DVM-input and leakage resistances were substantially reduced by using a proper measurement method. The method is being further investigated for measuring high-ohm standard resistors.
Keywords