DocumentCode
3801581
Title
Feasibility of Employing Model-Based Optimization of Pulse Amplitude and Electrode Distance for Effective Tumor Electropermeabilization
Author
Davorka Sel;Alenka Maek Lebar;Damijan Miklavcic
Author_Institution
Fac. of Electr. Eng., Ljubljana Univ.
Volume
54
Issue
5
fYear
2007
fDate
5/1/2007 12:00:00 AM
Firstpage
773
Lastpage
781
Abstract
In electrochemotherapy (ECT) electropermeabilization, parameters (pulse amplitude, electrode setup) need to be customized in order to expose the whole tumor to electric field intensities above permeabilizing threshold to achieve effective ECT. In this paper, we present a model-based optimization approach toward determination of optimal electropermeabilization parameters for effective ECT. The optimization is carried out by minimizing the difference between the permeabilization threshold and electric field intensities computed by finite element model in selected points of tumor. We examined the feasibility of model-based optimization of electropermeabilization parameters on a model geometry generated from computer tomography images, representing brain tissue with tumor. Continuous parameter subject to optimization was pulse amplitude. The distance between electrode pairs was optimized as a discrete parameter. Optimization also considered the pulse generator constraints on voltage and current. During optimization the two constraints were reached preventing the exposure of the entire volume of the tumor to electric field intensities above permeabilizing threshold. However, despite the fact that with the particular needle array holder and pulse generator the entire volume of the tumor was not permeabilized, the maximal extent of permeabilization for the particular case (electrodes, tissue) was determined with the proposed approach. Model-based optimization approach could also be used for electro-gene transfer, where electric field intensities should be distributed between permeabilizing threshold and irreversible threshold-the latter causing tissue necrosis. This can be obtained by adding constraints on maximum electric field intensity in optimization procedure
Keywords
"Electrodes","Neoplasms","Electrical capacitance tomography","Constraint optimization","Brain modeling","Solid modeling","Pulse generation","Finite element methods","Computational geometry","Voltage"
Journal_Title
IEEE Transactions on Biomedical Engineering
Publisher
ieee
ISSN
0018-9294
Type
jour
DOI
10.1109/TBME.2006.889196
Filename
4155015
Link To Document