  • Abstract
    This paper describes a X-rays microdetector based on an array of wells filled with scintillator crystals. A thick-film of aluminum (500 μm) is etched by Laser Ablation Technique (LAT) in order to achieve a square well with side of 100 μm and 490 μm deep vertical sidewalls. The aluminum thick-film was chosen because it is a good reflector in the visible range and for improving the number of photons collected by each photodetector. Also, the cross-talk between adjacent wells is reduced. As first approach, for a 4×4 wells array prototype, LAT is an inexpensive process when compared with Deep Reactive Ion Etching (DRIE) and avoids the need of etching masks. In this system the X-ray energy is first converted to visible light by the scintillator. This light is then detected by a photodetector fabricated in a standard CMOS process.
  • Keywords
    X-ray detection; arrays; biomedical equipment; diagnostic radiography; laser ablation; solid scintillation detectors; 100 micron; 490 micron; 500 micron; Al; aluminium thick-film; collected photons number improvement; cross-talk; deep reactive ion etching; maskless process; medical instrumentation; radiographic equipment; standard CMOS process; visible light; visible range reflector; wells array; Biomedical imaging; Brightness; Etching; Laser ablation; Laser theory; Optical arrays; Physics; Radiography; Radiology; X-ray lasers;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Sensors, 2002. Proceedings of IEEE
  • Print_ISBN
    0-7803-7454-1
  • Type

    conf

  • DOI
    10.1109/ICSENS.2002.1037158
  • Filename
    1037158