    MRI-compatible ultrasonic probe for minimally invasive therapy

  • Author

    Azuma, Takashi ; Sasaki, Kazuaki ; Kawabata, Ken-ichi ; Osada, Akiko ; Itagaki, Hiroyuki ; Komura, Kazumi ; Takahashi, Tetsuhiko ; Ishida, Kazunari ; Satoh, Yutaka ; Umemura, Shin-ichiro

  • Author_Institution
    Central Res. Lab., Hitachi Ltd., Kokubunji, Japan
  • Volume
    2
  • fYear
    2002
  • fDate
    8-11 Oct. 2002
  • Firstpage
    1465
  • Abstract
    The effects of magnetic resonance imaging (MRI)-compatible ultrasound (US) probes on MRI monitoring were evaluated. It was found that MRI-compatible US probes with backing material containing 100 ppm ferrite, did not disturb MR monitoring except within a few mm radius from the US probe´s position. MRI temperature monitoring of an excised swine liver irradiated with a high-intensity focused US beam from an MRI-compatible therapeutic transducer with US image guidance was then performed, and the potential usefulness of such a therapeutic system in minimally invasive therapy was demonstrated.
  • Keywords
    biomedical MRI; biomedical transducers; biomedical ultrasonics; liver; patient monitoring; probes; radiation therapy; surgery; temperature measurement; ultrasonic transducers; MR monitoring; MRI-compatible therapeutic transducer; MRI-compatible ultrasonic probe; US image guidance; backing material; excised swine liver; ferrite; high-intensity focused US beam; magnetic resonance monitoring; minimally invasive therapy; surgery time minimization; Ferrites; Magnetic materials; Magnetic resonance imaging; Medical treatment; Minimally invasive surgery; Monitoring; Probes; Temperature measurement; Temperature sensors; Ultrasonic imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium, 2002. Proceedings. 2002 IEEE
  • ISSN
    1051-0117
  • Print_ISBN
    0-7803-7582-3
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2002.1192573
  • Filename
    1192573