DocumentCode :
393406
Title :
Improvement in temperature scale realization above the copper point
Author :
Yamamoto, N. ; Yamada, Y.
Author_Institution :
Nat. Metrol. Inst. of Japan, Agency of Ind. Sci. & Technol., Tsukuba, Japan
Volume :
1
fYear :
2002
fDate :
5-7 Aug. 2002
Firstpage :
42
Abstract :
For improvement of the high-temperature scale realization, the spectral responsivity measurement uncertainty, which is the dominant source of uncertainty in the high temperature range, was investigated. A new single grating monochromator was introduced for this investigation, by means of which the target uncertainty of 0.03 nm was realized.
Keywords :
diffraction gratings; monochromators; spectral methods of temperature measurement; temperature scales; copper point; high temperature range; single grating monochromator; spectral responsivity measurement uncertainty; temperature scale realization; Copper; Filters; Gratings; Interference; Light sources; Metrology; Silver; Temperature distribution; Uncertainty; Wavelength measurement;