• DocumentCode
    393522
  • Title

    Development and evaluation of the tin point in the ITS-90

  • Author

    Sato, K. ; Tamba, J. ; Kishimoto, I. ; Arai, M.

  • Author_Institution
    Nat. Inst. of Adv. Ind. Sci. & Technol., Nat. Metrol. Inst. of Japan, Ibaraki, Japan
  • Volume
    2
  • fYear
    2002
  • fDate
    5-7 Aug. 2002
  • Firstpage
    1049
  • Abstract
    The freezing point of tin, one of the temperature fixed-points defined in the ITS-90, is newly developed and evaluated. The temperature stability of the furnace is within ±1.5 mK for 24 hours and temperature distribution is within 2 mK over 20 cm. To realize a stable plateau after a large amount of supercooling, the air blow cooling method is introduced instead of the conventional method. The temperature during the plateaus decreases only 0.3 mK for 8 hours and a good repeatability within ±0.026 mK is obtained in ten measurements.
  • Keywords
    furnaces; temperature control; temperature measurement; ITS-90; air blow cooling; calibration; furnace; repeatability; supercooling; temperature fixed point; temperature stability; tin point; Aluminum; Electrical resistance measurement; Furnaces; Metrology; Platinum; Stability; Temperature control; Temperature distribution; Thermal resistance; Tin;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    SICE 2002. Proceedings of the 41st SICE Annual Conference
  • Print_ISBN
    0-7803-7631-5
  • Type

    conf

  • DOI
    10.1109/SICE.2002.1195319
  • Filename
    1195319