Energy adjustment method based on transthoracic impedance for a biphasic defibrillator
Author
Portela, A.E. ; Folgueras, J. ; Colorado, O.
Author_Institution
Inst. Central de Investigacion Digital, Havana, Cuba
Volume
1
fYear
2003
fDate
17-21 Sept. 2003
Firstpage
155
Abstract
With the use of the biphasic truncated exponential waveform (BTE) for high energy discharges, a new concept was introduced for dealing with defibrillation: compensation (or adjustment) of the discharge voltage and time duration, as a function of transthoracic impedance. Transthoracic impedance is one of the parameters that determine the intensity and wave shape characteristics employed for the external defibrillation of a patient. A method for impedance compensation based on the impedance values measured both just before the discharge takes place and during the discharge across the patient´s thorax is proposed. The algorithm was tested by producing discharges in resistors simulating the patient and measuring the delivered energy. Results are presented and discussed, concluding that an excellent impedance compensation can be achieved.
Keywords
bioelectric phenomena; biomedical measurement; defibrillators; electric impedance measurement; patient diagnosis; patient treatment; biphasic defibrillator; defibrillation; energy adjustment method; high energy discharges; impedance compensation; thorax; transthoracic impedance; Capacitors; Cardiac arrest; Charge measurement; Current measurement; Defibrillation; Energy measurement; Impedance measurement; Shape; Thorax; Voltage;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 2003. Proceedings of the 25th Annual International Conference of the IEEE
ISSN
1094-687X
Print_ISBN
0-7803-7789-3
Type
conf
DOI
10.1109/IEMBS.2003.1279544
Filename
1279544
Link To Document