DocumentCode :
40895
Title :
A 0.6–107 µW Energy-Scalable Processor for Directly Analyzing Compressively-Sensed EEG
Author :
Shoaib, Mohammed ; Kyong Ho Lee ; Jha, Niraj K. ; Verma, Naveen
Author_Institution :
Dept. of Electr. Eng., Princeton Univ., Princeton, NJ, USA
Volume :
61
Issue :
4
fYear :
2014
fDate :
Apr-14
Firstpage :
1105
Lastpage :
1118
Abstract :
Compressive sensing has been used to overcome communication constraints (energy and bandwidth) in low-power sensors. In this work, we present a seizure-detection processor that directly uses compressively-sensed electroencephalograms (EEGs) for embedded signal analysis. In addition to addressing communication, this has two advantages for local computation. First, with compressive sensing, reconstruction costs are typically severe, precluding embedded analysis; directly analyzing the compressed signals circumvents reconstruction costs, enabling embedded analysis within applications. Second, compared to Nyquist-sampled signals, the use of compressed representations reduces the computational energy of signal analysis due to the reduced number of signal samples. We describe an algorithmic formulation as well as a hardware architecture that enables two strong power-management knobs, wherein application-level performance can scale with computational energy. The two knobs are parameterized as follows: 1) ξ, which quantifies the amount of data compression, and 2) ν, which determines the approximation error within the proposed compressed-domain processing algorithm. For ξ and ν in the range 2-24×, the energy to extract signal features (over 18 channels) is 70.8-1.3 nJ, and the detector´s performance for sensitivity, latency, and specificity is 96-91%, 4.7-5.3 sec., and 0.17-0.30 false-alarms/hr., respectively (compared to a baseline performance of 96%, 4.6 sec., and 0.15 false-alarms/hr.).
Keywords :
compressed sensing; data compression; electroencephalography; feature extraction; medical signal processing; sensitivity; signal reconstruction; signal sampling; Nyquist-sampled signals; algorithmic formulation; application-level performance; approximation error; communication constraints; compressed representations; compressed signals circumvents reconstruction costs; compressed-domain processing algorithm; compressively-sensed EEG; compressively-sensed electroencephalograms; computational energy; data compression; embedded signal analysis; energy-scalable processor; hardware architecture; local computation; low-power sensors; power 0.6 muW to 107 muW; power-management knobs; seizure-detection processor; sensitivity; signal feature extraction; Approximation methods; Detectors; Electroencephalography; Feature extraction; Integrated circuits; Signal analysis; Circuits; compressive sensing; energy efficiency; machine learning; seizure detection; signal processing;
fLanguage :
English
Journal_Title :
Circuits and Systems I: Regular Papers, IEEE Transactions on
Publisher :
ieee
ISSN :
1549-8328
Type :
jour
DOI :
10.1109/TCSI.2013.2285912
Filename :
6693771
Link To Document :
بازگشت