DocumentCode :
409572
Title :
Evaluation of arterial wall imaging with 16-slice multi-detector computed tomography
Author :
Chan, R.C. ; Ferencik, M. ; Wu, T. ; Hoffmann, U. ; Brady, T.J. ; Achenbach, S.
Author_Institution :
Dept. of Radiol., Harvard Med. Sch., Boston, MA, USA
fYear :
2003
fDate :
21-24 Sept. 2003
Firstpage :
661
Lastpage :
664
Abstract :
We have characterized the performance of 16-slice cardiac MDCT for assessment of coronary disease burden and lesion composition under optimal conditions. Morphology measurements with MDCT were accurate and reproducible. Detection of lipid-containing regions was possible with sensitivities ≥ 88% and specificity of 96% for regions ≥ 1.5 mm in size with ≥ 30% lipid. Estimation of lipid pool dimension is difficult and lipid pool CT densities cannot be used directly to determine the %lipid content due to noise and blurring. These results represent the best-case performance of MDCT and help to guide interpretation of images obtained in vivo.
Keywords :
blood vessels; cardiovascular system; computerised tomography; diseases; lipid bilayers; medical image processing; 16-slice multi-detector computed tomography; arterial wall imaging; coronary disease burden; lesion composition; lipid-containing region detection; Arteries; Cardiac disease; Cardiovascular diseases; Computed tomography; Imaging phantoms; In vivo; Lesions; Lipidomics; Morphology; Spatial resolution;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Computers in Cardiology, 2003
ISSN :
0276-6547
Print_ISBN :
0-7803-8170-X
Type :
conf
DOI :
10.1109/CIC.2003.1291242
Filename :
1291242