• DocumentCode
    410274
  • Title

    Fabrication and performance of 40-60 MHz annular arrays

  • Author

    Brown, Jeremy A. ; Morton, Christine E. ; Lockwood, Geoffrey R.

  • Author_Institution
    Queen´´s Univ., Kingston, Ont., Canada
  • Volume
    1
  • fYear
    2003
  • fDate
    5-8 Oct. 2003
  • Firstpage
    869
  • Abstract
    The fabrication and performance of miniature high frequency (40-60 MHz) annular arrays is described. The arrays were fabricated using photolithography to define the electrode pattern, and ultrasonic wire bonding to make electrical contacts. The transducers were 2 mm in diameter with seven equal area elements and a centre frequency of 50 MHz. A -6 dB two-way beam width of 75 microns was found at f/2. The radiation pattern decreased smoothly to less than -60 dB at a distance of 550 microns.
  • Keywords
    antenna radiation patterns; biomedical transducers; biomedical ultrasonics; electrical contacts; photolithography; ultrasonic bonding; ultrasonic transducer arrays; 2 mm; 40 to 60 MHz; 550 micron; annular array fabrication; electrical contacts; electrode pattern; photolithography; radiation pattern; transducers; two way beam; ultrasonic wire bonding; Costs; Dielectric substrates; Electrodes; Fabrication; Finite element methods; Focusing; Frequency; Machining; Phased arrays; Ultrasonic transducer arrays;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics, 2003 IEEE Symposium on
  • Print_ISBN
    0-7803-7922-5
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2003.1293537
  • Filename
    1293537