Water vapour partial pressure control in neonatal incubator using prevision by multiple models
Author
Amorim, M.F. ; Oliveira, G.H.C. ; Silva, D.S.R. ; Kiapuchinski, L.M.
Author_Institution
Technol. in Healthcare Graduate Program, Pontificia Univ. Catolica do Parana, Curitiba, Brazil
Volume
1
fYear
2004
fDate
1-5 Sept. 2004
Firstpage
2356
Lastpage
2359
Abstract
One of the phenomena that affect the water loss in newborn infants is the humidity. The evaporation is determined by the vapour partial pressure. As well as temperature control, the reduction of the vapour partial pressure oscillations has an important role in the establishment of thermal comfort for newborn infants in incubators. This work presents an algorithm based on multiple models to water vapour partial pressure control. The tests have been made in an acrylic prototype simulating a commercial incubator. The results shows that the implemented control algorithm presents oscillations minor to ±0.08 kPa.
Keywords
biomedical transducers; biothermics; humidity control; medical control systems; obstetrics; acrylic prototype; humidity control; neonatal incubator; newborn infants; partial pressure control; thermal comfort; vapour partial pressure oscillation; water vapour; Ducts; Humidity control; Humidity measurement; Instruments; Pediatrics; Performance evaluation; Pressure control; Pressure measurement; Temperature control; Thermal factors; Humidity control; incubator; vapour partial pressure;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 2004. IEMBS '04. 26th Annual International Conference of the IEEE
Conference_Location
San Francisco, CA
Print_ISBN
0-7803-8439-3
Type
conf
DOI
10.1109/IEMBS.2004.1403683
Filename
1403683
Link To Document