DocumentCode
429437
Title
Electrical detection of the contact between a microinjection pipette and cells
Author
Lukkari, M.J. ; Karjalainen, M.I. ; Sarkanen, R. ; Linne, M.-L. ; Jalonen, T.O. ; Kallio, P.J.
Author_Institution
Inst. of Autom. & Control, Tampere Univ. of Technol., Finland
Volume
1
fYear
2004
fDate
1-5 Sept. 2004
Firstpage
2557
Lastpage
2560
Abstract
This work presents the basic design and tests of a device designed for detecting the contact between a microinjection pipette and cell membrane. The device facilitates the automation of the microinjection procedure of living adherent cells. The measurement of the contact is based on measuring the resistance of the pipette. Breakage and clogging of the pipette can be detected with the same measurement.
Keywords
bioelectric phenomena; biomedical equipment; biomedical measurement; biomembranes; cellular biophysics; electrical contacts; cell membrane; electrical contact detection; living adherent cells; microinjection pipette; micromanipulator; pipette breakage; pipette clogging; resistance measurement; Clamps; Contacts; Current measurement; Electrical resistance measurement; Electrodes; Immune system; Microinjection; Micromanipulators; Pulse measurements; Voltage; Contact; microinjection; micromanipulator;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 2004. IEMBS '04. 26th Annual International Conference of the IEEE
Conference_Location
San Francisco, CA
Print_ISBN
0-7803-8439-3
Type
conf
DOI
10.1109/IEMBS.2004.1403735
Filename
1403735