    The resonance frequencies of individual SonoVue™ contrast agent bubbles were measured optically by recording the radius-time curves of a single microbubble at 24 different frequencies. For these experiments, the Brandaris 128 fast framing camera was operated in a special segmented mode. The resonance frequencies found for SonoVue™ microbubbles are in good agreement with the modified Herring model for coated bubbles indicating that the shell only slightly affects the resonance frequency of this class of contrast bubbles.
  • Keywords
    biomedical ultrasonics; bubbles; resonance; ultrasonic measurement; Brandaris 128 fast framing camera; SonoVue contrast agent bubbles; coated bubbles; high-speed optical imaging; modified Herring model; radius-time curves; resonance frequency; ultrasound contrast agent; Cameras; Elasticity; Equations; Optical imaging; Optical recording; Optical scattering; Resonance; Resonant frequency; Surface tension; Ultrasonic imaging;